Anaesthetic Equipments – Berovan https://berovan.com Reliable distributor of the top-of-the-line medical supplies and equipments Wed, 21 May 2025 09:02:05 +0000 en-US hourly 1 MS Anesthesia Machine Only Dble SL210 https://berovan.com/product/ms-anesthesia-machine-only-dble-sl210/ Tue, 28 Nov 2017 18:45:55 +0000 http://www.berovan.com/product/ms-anesthesia-machine-only-dble-sl210/

Oxygen Ratio Controller

O2 and N2O needles are mutually controlled by a special interlocking gear to ensure the minimum O2 concentration for a fail-safe mechanism.

Each part of a respiratory circuit is movable. The direction of the attachment part of the breathing bag, and the expiration and inhalation port can be freely set up, and workability is remarkably improved.]]>

Oxygen Ratio Controller

O2 and N2O needles are mutually controlled by a special interlocking gear to ensure the minimum O2 concentration for a fail-safe mechanism.

Each part of a respiratory circuit is movable. The direction of the attachment part of the breathing bag, and the expiration and inhalation port can be freely set up, and workability is remarkably improved.]]>
MS Anesthesia Machine W/ 2 Vaporizer Aeon 8600A https://berovan.com/product/ms-anesthesia-machine-w-2-vaporizer-aeon-8600a/ Tue, 28 Nov 2017 18:48:44 +0000 http://www.berovan.com/product/ms-anesthesia-machine-w-2-vaporizer-aeon-8600a/ ✓ Optional functions monitoring O2, CO2, and anesthetic gas
✓ Flowmeter provides minimum 25% concentration of oxygen in O2/N2O mixture
✓ Absorb system sterilized with high temperature and high pressure
✓ Enhances usability with auxiliary oxygen, suction, auxiliary power source, monitor arm, infusion pump etc. ]]>
✓ Optional functions monitoring O2, CO2, and anesthetic gas
✓ Flowmeter provides minimum 25% concentration of oxygen in O2/N2O mixture
✓ Absorb system sterilized with high temperature and high pressure
✓ Enhances usability with auxiliary oxygen, suction, auxiliary power source, monitor arm, infusion pump etc. ]]>
MS Anesthesia Machine W/2 Vaporizer ORICA9600 https://berovan.com/product/ms-anesthesia-machine-w2-vaporizer-orica9600/ Tue, 28 Nov 2017 18:39:40 +0000 http://www.berovan.com/product/ms-anesthesia-machine-w2-vaporizer-orica9600/
It has a user-friendly design, incorporates new technology and provides safe and effective treatment options for the clinician. A9600 includes Adult and Child modes that provide patient-appropriate defaults and ranges.

✓ Advanced features for therapy delivery are also easy to use. The single-turn APL valve on A9600 includes a quick-release function to quickly lower patient breathing pressure and accurately set pressure limit
✓ Automatic Compliance compensation along with Fresh Gas Flow compensation help clinician to deliver accurate and precise ventilation therapy
✓ Full waveform display including integrated Spirometry provides loops when desired for improved clinical data analysis
✓ System provides a minimum of 21% O2 concentration at all times utilizing a pneumatic Oxygen Ratio Controller. This enhances patient safety over systems that utilize electronic or software controlled ORC functions
✓ Adjustable table surface lighting system (2-levels of intensity) combined with a spotlight is suitable for use in low light conditions
✓ Impressive array of standard features improves the system usability: auxiliary oxygen flow meter and auxiliary AC power outlets
✓ Breathing system is heated to reduce condensation
✓ Absorber is compatible with Standard Prepaks or loose-fill CO2 Absorbent
✓ “GCX-type” rails support easy mounting of other devices to the workstation
✓ 6 traditional Gas Supply Pressure Gauges allow for easy status monitoring of hospital wall gas supplies and gas cylinders/tanks
✓ 3 locking drawers for storage
✓ Standard Selectatec™-compatible mounts hold two vaporizers]]>
A9600 Anesthesia Workstation Technical Parameters

Ventilation Modes

  • Volume Control – Intermittent Positive Pressure Ventilation (IPPV)
  • Pressure Control Ventilation (PCV)
  • Synchronized Intermittent Mandatory Ventilation with Pressure Support (SIMV + PS)
  • Pressure Support with Apnea backup (PS)
  • Manual

Parameters and Ranges

  • Pressure target: 5 – 70 cmH2O
  • Pressure support (delta P): 3 – 50 cmH2O
  • Tidal Volume: 20 – 1500 mL
  • Breathing Freq.: 4-100 bpm
    2-60 bpm in PS, SIMV
  • TINSP: 0.2-5.0s
  • PEEP: OFF, 3 – 30 cmH2O
  • FreqMIN: 2 – 60 bpm
  • T pause: OFF, 5%-60%
  • Trigger: 1 – 15 L/min
  • I:E Ratio: 4:1-1:8
  • TSLOPE: 0 – 2 s
  • Vaporizers: Sevoflurane, Desflurane, Isoflurane, Halothane, Enflurane

Monitoring
Continuous monitoring of inspiratory O2 concentration, breathing frequency, airway pressure (Paw, Peak, Pmean, PEEP), minute volume and tidal volume. The measured parameters are displayed
as large, easy to read digital values. Airway pressure, flow, and CO2 (optional) are shown in graphical waveforms. Optional Gas Monitoring for CO2 and 5 Anesthetic Agents with Agent Identification

]]>
MS Anesthesia Machine W/2 Vaporizer ORICA9800 https://berovan.com/product/ms-anesthesia-machine-w2-vaporizer-orica9800/ Tue, 28 Nov 2017 18:43:05 +0000 http://www.berovan.com/product/ms-anesthesia-machine-w2-vaporizer-orica9800/
It has a user-friendly design, incorporates new technology and provides safe and effective treatment options for the clinician.
A9800 provides complete anesthesia ventilation capabilities that include traditional and “intensive-care type” ventilation modes like PCV-VG (volume guarantee). Low-flow anesthesia delivery creates savings by lowering facility gas usage. Sophisticated ventilation capabilities of A9800 will meet the needs of the full patient range. The integrated Electronic flowmeter provides accurate monitoring and intuitive operation.
]]>
Ventilation Modes
· Volume Controlled Ventilation (VCV)
· Pressure Controlled Ventilation (PVC)
· Synchronized Intermittent Mandatory Ventilation w / Pressure Support (SIMV + PS) with Volume or Pressure-type mandatory breaths
· Pressure Support (PS) with Apnea Backup
· Pressure Controlled Ventilation with Volume Guarantee (PCV-VG)
· Manual Ventilation

Parameters and Ranges
· Pressure target: 5 – 70 cmH2O
· Pressure support (delta P): 3 – 50 cmH2O
· Tidal Volume: 20 – 1500 mL
· Breathing Freq.: 2 – 100 bpm
  2 – 60 bpm in PS
  2 – 100 bpm in SIMV-VC and SIMV-PC
  4 – 100 bpm in other modes
· TINSP: 0.2 – 5.0 s
· PEEP: OFF, 3 – 30 cmH2O
· FreqMIN: 2 – 60 bpm
· Tpause: OFF, 5% – 60%
· Trigger: 1 – 15 L/min
· I:E Ratio: 4:1 – 1:8
· TSLOPE: 0 – 2 s
· Vaporizers: Sevoflurane, Desflurane, Isoflurane, Halothane, Enflurane

Monitoring
· Continous monitoring of inspiratory O2 concentration, breathing frequency, airway pressure (Ppeak, Plat, Pmean, PEEP), minute volume and tidal volume.
· The measured parameters are displayed as large, easy to read digital values. Airway pressure, flow, volume and CO2 (optional) are shown in graphical waveforms.

]]>